General Instrument Setup
R&S
®
FSW
679User Manual 1173.9411.02 ─ 43
Name...........................................................................................................................679
Unit..............................................................................................................................679
Compatibility................................................................................................................679
Activating / Deactivating..............................................................................................679
Comment.....................................................................................................................679
Included Transducer Lines in Overview (View Filter)..................................................680
Adjust Ref Level..........................................................................................................680
Create New Line......................................................................................................... 680
Edit Line...................................................................................................................... 680
Copy Line....................................................................................................................680
Delete Line..................................................................................................................680
Name
The name of the stored transducer line.
Unit
The unit in which the y-values of the data points of the transducer line are defined.
The following units are available:
●
dB
●
dBm
●
dBmV
●
dBμV
●
dBμV/m
●
dBμA
●
dBμA/m
●
dBpW
●
dBpT
Additional units available only for installed R&S FSW-K54 (EMI measurements) option:
●
dBmV/MHz (normalized to 1 MHz)
●
dBµV/MHz (normalized to 1 MHz)
●
dBµA/MHz (normalized to 1 MHz)
Compatibility
Indicates whether the transducer factors are compatible with the current measurement
settings.
For more information on which conditions a transducer line must fulfill to be compati-
ble, see Chapter 12.3.1, "Basics on Transducer Factors", on page 675.
Activating / Deactivating
Activates/deactivates the transducer line. Up to 8 transducer lines can be active at the
same time.
Remote command:
[SENSe:]CORRection:TRANsducer:SELect on page 1270
[SENSe:]CORRection:TRANsducer[:STATe] on page 1270
Comment
An optional description of the transducer line.
Transducers