General Instrument Setup
R&S
®
FSW
680User Manual 1173.9411.02 ─ 43
Included Transducer Lines in Overview (View Filter)
Defines which of the stored transducer lines are included in the overview. The view can
be restricted to compatible transducer lines only or include all transducer lines found.
Whether a line is compatible or not is indicated in the Compatibility setting.
Adjust Ref Level
Activates or deactivates the automatic adjustment of the reference level to the selected
transducer factor.
"Auto"
Activates the automatic adjustment. The original dynamic range is
restored by shifting the reference level by the maximum transducer
factor.
"Manual"
Deactivates the automatic adjustment. Adjust the reference level via
the "Amplitude" menu.
Remote command:
[SENSe:]CORRection:TRANsducer:ADJust:RLEVel[:STATe] on page 1268
Create New Line
Create a new transducer line.
Remote command:
[SENSe:]CORRection:TRANsducer:SELect on page 1270
Edit Line
Edit an existing transducer line configuration.
Copy Line
Copy the selected transducer line configuration to create a new line.
Delete Line
Delete the selected transducer line.
Remote command:
[SENSe:]CORRection:TRANsducer:DELete on page 1269
12.3.2.2 Transducer Factors
Access: [Setup] > "Transducer" > "Edit Line" / "Copy Line" / "New Line"
The settings and functions available for individual transducer lines are described here.
For instructions on creating and editing transducer lines see Chapter 12.3.4, "How to
Configure the Transducer", on page 684.
Transducers