General Instrument Setup
R&S
®
FSW
693User Manual 1173.9411.02 ─ 43
An active external preamplifier is also included in the calculation of the combined user-
defined frequency response correction filter and displayed in the preview for SnP files
(see "Preview" on page 696).
When disabled, no compensation is performed even if an external preamplifier remains
connected.
Remote command:
INPut<ip>:EGAin[:STATe] on page 1044
Apply to
Determines which channels the correction settings are applied to.
Note: In MSRA mode, the settings are always applied to all active measurement chan-
nels.
"All"
The frequency response correction settings are applied to all active
measurement channels with the same input type. A new filter is cal-
culated for each measurement channel as soon as you switch to it, or
when a measurement is performed (e.g. by the Sequencer).
"Channel"
The frequency response correction settings are applied to the cur-
rently selected channel only.
Remote command:
[SENSe:]CORRection:FRESponse<si>:USER:SCOPe on page 1279
Load Settings
Loads a stored saveset for a user-defined frequency response correction scenario.
Existing settings in the dialog box are overwritten. The settings apply to the currently
selected channel or all channels, depending on the Apply to setting.
Only .dfl files can be loaded. The default storage directory for correction files is
C:\R_S\INSTR\USER\FResponse.
Remote command:
[SENSe:]CORRection:FRESponse<si>:USER:LOAD on page 1279
Save Settings
Stores a saveset for a user-defined frequency response correction scenario. As for all
instrument settings, a .dfl file is created for the correction data. Note that only the
settings defined in this dialog box are stored, not the contents of the files themselves.
Whether the settings for the currently selected channel only or for all channels are
stored depends on the Apply to setting.
The default storage directory is C:\R_S\INSTR\USER\FResponse.
Remote command:
[SENSe:]CORRection:FRESponse<si>:USER:STORe on page 1287
Clear Settings
Clears all current user-defined frequency response correction settings, either for the
currently selected channel only or for all channels, depending on the Apply to setting.
Merely files specific to the R&S FSW itself remain in the list of Touchstone files.
Remote command:
[SENSe:]CORRection:FRESponse<si>:USER:PRESet on page 1279
Frequency Response Correction (R&S
FSW-K544)