Manual Programming with Guardrails™ Suite MX
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 89
3. Prime (see Getting started on page 62).
4. Press the CHANNEL SELECT key.
Ensure that the displayed syringe manufacturer and syringe size correctly identify the installed
syringe. Mismatches might cause an under-infusion or over-infusion to the patient that could
result in serious injury and/or death. For a list of compatible syringes, refer to Compatible
Syringes (Syringe Module) on page 175.
pump module syringe module