Manual Programming with Guardrails™ Suite MX
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 97
• If the programmed continuous dose infusion is outside the soft limit for that care area, an audio
alert sounds and a visual prompt appears before programming can continue. If
Yes soft key is
pressed, programming continues; if
No soft key is pressed, infusion needs to be reprogrammed.
• If the programmed continuous dose infusion is outside the hard limit for that care area, an audio
alert sounds and a visual prompt appears before programming can continue. Infusion needs to be
reprogrammed.
• If a dose outside of the soft limits has been entered and verified as correct, message display also
shows either “LLL” for a low dose or “
↑↑↑
” for a high dose.
• If a soft limit is overridden,
G icon is displayed. When G soft key is pressed, all applicable out-of-
range limits are listed.
VOLUME
INFUSED
AUDIO
ADJUST
Midtown Hospital
Adult ICU
A
VTBI = 250 mL
G