Manual Programming with Guardrails™ Suite MX
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 97
If the programmed continuous dose infusion is outside the soft limit for that care area, an audio
alert sounds and a visual prompt appears before programming can continue. If
Yes soft key is
pressed, programming continues; if
No soft key is pressed, infusion needs to be reprogrammed.
If the programmed continuous dose infusion is outside the hard limit for that care area, an audio
alert sounds and a visual prompt appears before programming can continue. Infusion needs to be
reprogrammed.
If a dose outside of the soft limits has been entered and verified as correct, message display also
shows either “LLL” for a low dose or “
↑↑↑
” for a high dose.
If a soft limit is overridden,
G icon is displayed. When G soft key is pressed, all applicable out-of-
range limits are listed.
VOLUME
INFUSED
AUDIO
ADJUST
Midtown Hospital
Adult ICU
A
VTBI = 250 mL
G