• Syringe module:
°
If ALL Mode is enabled for syringe configuration in data set, ALL is displayed in VTBI field and
estimated available volume in syringe is displayed.
or
°
If ALL Mode is disabled for syringe configuration in data set, VTBI ALL option is not available
and estimated available volume in syringe is displayed when
VTBI soft key is pressed.
°
To enter or change a numeric VTBI value, press the VTBI soft key and use numeric data entry
keys.
°
To deliver entire contents of syringe: Keep an ALL VTBI value, or press the ALL soft key to
change a numeric
VTBI value to ALL.
3. To enter volume duration, press the
DURATION soft key and use numeric data entry keys.
Rate is automatically calculated.
4. Verify correct infusion parameter entry and press the START soft key.
RATE =
VTBI
DURATION
>Enter Duration
RATE
VOLUME
A
TPN
Guardrails Fluid Setup
PRIMARY INFUSION
24 00h min
START
41.7 mL/h
1000
mL
PAUSE