Chapter 3—BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
180 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
Feature Definition—Syringe Module
NEOI Snooze
Optional capability to remind users NEOI alarm has been silenced. This is shared
syringe and PCA setting.
Numeric Pressure Display
Dynamic pressure display is numerical and only available when pressure sensing
disc is inserted.
Occlusion Pressure
A complete range of downstream occlusion detection options is provided.
• With pressure sensing disc: Downstream o
cclusion alarm threshold is selectable
between 25 and 1000 mmHg, in 1 mmHg increments.
•
Without pressure sensing disc: Downstream occlusion alarm threshold
can be
set to low, medium, or high.
Pressure Sensing Disc
When installed, pressure sensing disc significantly improves instrument’s pressure
sensing capabilities for a faster occlusion detection time, and makes following
features available:
• Auto Pressure
• Back-Off
• Customizable Pressure Alarm Settings (see Occlusion Pressure on
page 180)
•Fast Start
• Pressure Tracking
Priming
Allows a limited volume of fluid to be delivered in order to prime administration set
prior to being connected to a patient or after changing a syringe. When priming, a
single continuous press of
PRIME
soft key delivers up to 2 mL of priming fluid.
Syringe Empty
Instrument gives an alert and stops when an empty syringe is detected.
Syringe Volume Detection
System automatically detects fluid volume in a syringe when it is inserted.