General Information
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 193
Environmental
Conditions:
Equipment
Orientation:
To ensure proper operation, the system must remain in an upright position.
Flow Rate
Programming:
Flow rate range is from 0.01 to 999 mL/h and can be selected as follows:
Rate restriction by syringe size:
Fluid Ingress
Protection:
IPX2, Drip Proof
Infusion
Pressure,
Maximum:
Without Pressure Sensing Disc: approximately 800 mmHg (actual occlusion pressure varies
based on syringe size and manufacturer)
With Pressure Sensing Disc: 1060 mmHg
KVO (Keep
Vein Ope n)
Rate:
Factory default setting is 1 mL/h if set rate is 1 mL/h or above; or set rate, if rate is 0.9 mL/h or
below.
Symbols Meaning Operating Storage/Transport
Atmospheric Pressure 525 - 4560 mmHg
(70 - 608 kPa)
375 - 760 mmHg
(50 - 101 kPa)
Relative Humidity
(Avoid prolonged
exposure to relative
humidity >90%)
20 - 90%
Noncondensing
5 - 90%
Noncondensing
Temperature Range 41 - 104°F
(5 - 40°C)
-4 - 140°F
(-20 - 60°C)
Flow Rates (mL) Selectable Increments (mL/h)
0.01 - 9.99 0.01
10 - 99.9 0.1
100 - 999 1
Syringe Size (mL) Flow Rate Range (mL/h)
50/60 0.1 - 999
30 0.1 - 650
20 0.1 - 500
10 0.1 - 250
5 0.1 - 150
30.01 - 100
1 0.01 - 30