EasyManua.ls Logo

BD Alaris - Page 91

Default Icon
376 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Getting started
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 77
7. Insert pressure sensing disc (if used), as follows:
a. Orient pressure sensing disc, as follows:
°
fluid side up (patient side down)
°
cavity forward (membrane toward instrument)
b. Gently slide pressure sensing disc up
into slot in pressure sensing disc housing.
WARNING
When the pressure sensing disc is not being used and an occlusion occurs, there is a risk of
infusing pressurized buildup of infusates upon correction of the occlusion. To avoid an
inadvertent bolus, relieve the pressure before restarting the infusion.
Syringe Module
Slot in Pressure
Sensing Disc
Housing
Pressure
Sensing Disc
Patient Side
Fluid Side

Table of Contents

Other manuals for BD Alaris

Related product manuals