Section
3: Functions and features Model 2450 Interactive SourceMeter® Instrument
Reference Manual
3-28 2450-901-01 Rev. B/September 2013
Storing data in buffers
Using SCPI commands:
The following SCPI commands are associated with data storage operation:
Command Description
:TRACe:MAKE
This function creates a user-defined reading buffer. You cannot use
this command on the default buffers. See Creating buffers (on page
3-12). Also see :TRACe:MAKE (on page 6-114).
:TRACe:SAVE
This function saves data from the specified reading buffer to a USB
flash drive. See Storing readings in buffers (on page 3-17
). Also see
:TRACe:SAVE (on page 6-116).
:TRACE:SAVE:APPend
This function appends data from the reading buffer to a file on the
USB flash drive. See :TRACE:SAVE:APPend (on page 6-118).
:TRACe:POINts
This command reads the number of readings a buffer can store. This
allows you to change the number of readings the buffer can store.
See :TRACe:POINts (on page 6-115).
:TRACe:CLEar
This command clears all readings and statistics from the specified
buffer. See Clearing buffers (on page 3-26). Also see :TRACe:CLEar
(on page 6-108).
:TRACe:STATistics:CLEar
This command clears the statistical information associated with the
specified buffer. This command does not clear the readings.
:TRACe:FILL:MODE
This command determines if a reading buffer is filled continuously or
is filled once and stops. See :TRACe:FILL:MODE (on page 6-112).
:TRACe:LOG:STATe
This command indicates whether the reading buffer should log
informational events. See :TRACe:LOG:STATe (on page 6-113).
:TRACe:ACTual?
This command contains the number of readings in the specified
buffer. See :TRACe:ACTual? (on page 6-107)
:TRACe:DELete
This command deletes a buffer. See :TRACe:DELete (on page 6-
111).

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm