D: Model 2450 in a Model 2400 application Model 2450 Interactive SourceMeter® Instrument Referenc
D-8 2450-901-01 Rev. B/September 2013
Model 2400 command
:CALCulate2:LIMit[1]:COMPliance:FAIL
:CALCulate2:LIMit[1]:COMPliance:FAIL?
Model 2450 command
Notes
Model 2400 command
:CALCulate2:LIMit[1]:COMPliance:SOURce2
:CALCulate2:LIMit[1]:COMPliance:SOURce2?
Model 2450 command
Notes
Model 2400 command
:CALCulate2:LIMit[1]:FAIL?
Model 2450 command
Notes
Model 2400 command
:CALCulate2:LIMit[1]:STATe
:CALCulate2:LIMit[1]:STATe?
Model 2450 command
Not available
Notes
To disable source limits, set the limit value to the maximum allowed by the
Model 2400 command
:CALCulate2:LIMit<x>:FAIL?
Model 2450 command
:CALCulate2:<function>:LIMit<y>:FAIL? (on page 6-15), where <function> is
VOLTage[:DC], CURRent[:DC], or RESistance
Notes
Note that :CALCulate2:LIMit4:FAIL? is not supported (related to contact
check, which is not provided on Model 2450).
Note that this only available for two limits in the Model 2450.
Model 2400 command
:CALCulate2:LIMit<x>:LOWer[:DATA]
:CALCulate2:LIMit<x>:LOWer[:DATA]?
Model 2450 command
:CALCulate2:<function>:LIMit<y>:LOWer[:DATA] (on page 6-16
), where
is
,
, or
Notes
Note that this only available for two limits in the Model 2450.
Model 2400 command
:CALCulate2:LIMit<x>:LOWer:SOURce2
:CALCulate2:LIMit<x>:LOWer:SOURce2?
Model 2450 command
Notes
Model 2400 command
:CALCulate2:LIMit<x>:PASS:SOURce2
:CALCulate2:LIMit<x>:PASS:SOURce2?
Model 2450 command
Not available
Notes
Model 2400 command
:CALCulate2:LIMit<x>:STATe
:CALCulate2:LIMit<x>:STATe?
Model 2450 command
:CALCulate2:<function>:LIMit<y>:STATe (on page 6-17), where <function>
is
,
, or
and
is
or
.
Notes
Note that this only available for two limits in the Model 2450.
To disable source limits, set the limit value to the maximum allowed by the