Appendix
D: Model 2450 in a Model 2400 application Model 2450 Interactive SourceMeter® Instrument
Reference Manual
D-12 2450-901-01 Rev. B/September 2013
Model 2400 command
:OUTPut[1]:ENABle:TRIPped?
Model 2450 command
Not available
Notes
Model 2400 command
OUTPut[1]:INTerlock:TRIPped?
Model 2450 command
:OUTPut[1]:INTerlock:TRIPped? (on page 6-33)
Notes
Model 2400 command
:OUTPut[1]:SMODe
:OUTPut[1]:SMODe?
Model 2450 command
:OUTPut[1]:<function>:SMODe (on page 6-31
), where
<function>
is
VOLTage[:DC] or CURRent[:DC]
OUTPut[1]:<function>:SMODe?
Notes
Model 2400 command
:OUTPut[1][:STATe]
:OUTPut[1][:STATe]?
Model 2450 command
:OUTPut[1][:STATe] (on page 6-34
)
:OUTPut[1][:STATe]?
Notes
READ?
Model 2400 command
:READ?
Model 2450 command
:READ? (on page 6-6)
Notes
Model 2450 allows you to choose different buffers and which buffer elements to
access from the buffers when you send the command.
ROUTe subsystem
Model 2400 command
:ROUTe:TERMinals
:ROUTe:TERMinals?
Model 2450 command
:ROUTe:TERMinals (on page 6-35)
:ROUTe:TERMinals?
Notes
SENSe subsystem
Model 2400 command
:MEASure:CURRent[:DC]?
:MEASure:RESistance?
:MEASure:VOLTage[:DC]?
:MEASure[:<function>]?
Model 2450 command
:MEASure:<function>? (on page 6-4
), where <function> is VOLTage[:DC],
CURRent[:DC]
, or
RESistance
.
Notes
Can specify a buffer in Model 2450 and which buffer elements to access from
the buffer when you specify the command. In the Model 2450, this command
changes the measurement function to the function in the command if it is not
already active, makes readings, and stores them in a reading buffer, which you
can specify. When it changes to that function, it recalls the settings as they
were the last time that function was active. It does not go to factory default
settings for the function (which the Series 2400 did). Also, in the Model 2450,
this command does not map to
CONFigure - READ - FETCh
.