Model 2450
Interactive SourceMeter® Instrument Reference Manual Appendix D:
Model 2450 in a Model 2400 application
2450-901-01 Rev. B/September 2013 D-23
Model 2400 command
:STATus:MEASurement:ENABle
:STATus:MEASurement:ENABle?
Model 2450 command
Not available
Notes
Use the questionable register to emulate this register; see the STATus
subsystem (on page 6-89) for information.
Model 2400 command
:STATus:MEASurement[:EVENt]?
Model 2450 command
Not available
Notes
Use the questionable register to emulate this register; see the STATus
subsystem (on page 6-89) for information.
Model 2400 command
:STATus:OPERation:CONDition?
Model 2450 command
:STATus:OPERation:CONDition? (on page 6-89)
Notes
In the Model 2450, you need to map events into the register (there are no set
bits). See Status model (on page C-1) for information.
Model 2400 command
:STATus:OPERation:ENABle
:STATus:OPERation:ENABle?
Model 2450 command
:STATus:OPERation:ENABle (on page 6-89)
:STATus:OPERation:ENABle?
Notes
In the Model 2450, you need to map events into the register (there are no set
bits). See Status model (on page C-1) for information.
Model 2400 command
:STATus:OPERation[:EVENt]?
Model 2450 command
:STATus:OPERation[:EVENt]? (on page 6-90)
Notes
In the Model 2450, you need to map events into the register (there are no set
bits). See Status model (on page C-1) for information.
Model 2400 command
:STATus:PRESet
Model 2450 command
:STATus:PRESet (on page 6-92)
Notes
Model 2400 command
:STATus:QUEStionable:CONDition?
Model 2450 command
:STATus:QUEStionable:CONDition? (on page 6-92)
Notes
In the Model 2450, you need to map events into the register (there are no set
bits). See Status model (on page C-1) for information.
Model 2400 command
:STATus:QUEStionable:ENABle
:STATus:QUEStionable:ENABle?
Model 2450 command
:STATus:QUEStionable:ENABle (on page 6-93)
:STATus:QUEStionable:ENABle?
Notes
In the Model 2450, you need to map events into the register (there are no set
bits). See Status model (on page C-1) for information.
Model 2400 command
:STATus:QUEStionable[:EVENt]?
Model 2450 command
:STATus:QUEStionable[:EVENt]? (on page 6-93)
Notes
In the Model 2450, you need to map events into the register (there are no set
bits). See Status model (on page C-1) for information.
Model 2400 command
:STATus:QUEue:CLEar
Model 2450 command
Not available
Notes
Use the event log; see Using the event log (on page 2-117).

Table of Contents