Interactive SourceMeter® Instrument Reference Manual Appendix D:
Model 2450 in a Model 2400 application
2450-901-01 Rev. B/September 2013 D-25
Model 2400 command
:SYSTem:CCHeck
:SYSTem:CCHeck?
Model 2450 command
Notes
No contact check on Model 2450.
Model 2400 command
:SYSTem:CCHeck:RESistance
:SYSTem:CCHeck:RESistance?
Model 2450 command
Notes
No contact check on Model 2450.
Model 2400 command
Model 2450 command
:SYSTem:CLEar (on page 6-96)
Notes
Model 2400 command
Model 2450 command
Notes
See Using the event log (on page 2-117) for changes to error reporting.
Model 2400 command
Model 2450 command
Not available
Notes
See Using the event log (on page 2-117) for changes to error reporting.
Model 2400 command
:SYSTem:ERRor:CODE[:NEXT]?
Model 2450 command
:SYSTem:ERRor:CODE[:NEXT]? (on page 6-99)
Notes
Returns only errors from the event log. See Using the event log (on page 2-
117) for changes to error reporting.
Model 2400 command
Model 2450 command
:SYSTem:ERRor:COUNt? (on page 6-99)
Notes
Returns only errors from the event log. See Using the event log (on page 2-
117) for changes to error reporting.
Model 2400 command
Model 2450 command
:SYSTem:ERRor[:NEXT]? (on page 6-98)
Notes
Returns only errors from the event log. See Using the event log (on page 2-
117) for changes to error reporting.
Model 2400 command
:SYSTem:GUARd
:SYSTem:GUARd?
Model 2450 command
Not available
Notes
Cable guard is the only option available on Model 2450.
Model 2400 command
Model 2450 command
Notes