Appendix
D: Model 2450 in a Model 2400 application Model 2450 Interactive SourceMeter® Instrument
Reference Manual
D-26 2450-901-01 Rev. B/September 2013
Model 2400 command
:SYSTem:LFRequency
:SYSTem:LFRequency?
Model 2450 command
:SYSTem:LFRequency? (on page 6-104)
Notes
Line frequency is always automatically detected in the Model 2450.
Model 2400 command
:SYSTem:LFRequency:AUTO
:SYSTem:LFRequency:AUTO?
Model 2450 command
Not available
Notes
Line frequency is always automatically detected in the Model 2450.
Model 2400 command
:SYSTem:LOCal
Model 2450 command
Not available
Notes
No RS-232 communications available in the Model 2450.
Model 2400 command
:SYSTem:MEMory:INITialize
Model 2450 command
Not available
Notes
No battery-backed RAM in the Model 2450.
Model 2400 command
:SYSTem:MEP:HOLDoff
Model 2450 command
Notes
Model 2400 command
:SYSTem:MEP[:STATe]?
Model 2450 command
Notes
Model 2400 command
:SYSTem:POSetup
:SYSTem:POSetup?
Model 2450 command
:SYSTem:POSetup (on page 6-105)
:SYSTem:POSetup?
Notes
Model 2400 command
:SYSTem:PRESet
Model 2450 command
Not available
Notes
Model 2400 command
:SYSTem:RCMode
:SYSTem:RCMode?
Model 2450 command
Not available
Notes
Model 2400 command
:SYSTem:RSENse
:SYSTem:RSENse?
Model 2450 command
[:SENSe[1]]:<function>:RSENse (on page 6-58), where <function> is
VOLTage[:DC] or RESistance.
[:SENSe[1]]:<function>:RSENse?
Notes