No. 99MBB122A
6-10
■ Operating procedure (Refer to “■ Accessing the Calibration Menu screen” in Section 6.2.)
Home screen to Main Menu
1 Check the calibration conditions on the Calibration Measurement
screen.
When the conditions are different from the precision roughness
specimen, modify the calibration conditions.
If the calibration conditions do not require modification, proceed to the
next step.
TIP
• For information about the procedure used to modify the
calibration conditions, refer to 6.4, “Setting the Nominal
Value of the Precision Roughness Specimen” and 6.5,
“Setting Calibration Conditions”.
2 Press the [START/STOP] key.
Start calibration measurement.
A calibration measurement with the precision roughness specimen
is performed, and Measurement Waveform Display screen is
displayed during the calibration measurement (while the detector is
traversing).
When the calibration measurement has been completed, the
measured value is displayed in the lower column.
3 Press the “Update” ([Red] key) to update the calibration value.
TIP
• To cancel the measured value, press the “Cancel” ([Blue]
key).
The calibration factor is updated.
Measurement Waveform Display
screen
Calibration Measurement screen
Calibration Measurement screen
Calibration Measurement screen