No. 99MBB122A
6-22
6.5.6 Setting the evaluation length to an arbitrary length
Set the evaluation length to an arbitrary length according to the precision roughness
specimen.
■ Operating procedure (Refer to “■ Accessing the Calibration Menu screen” in Section 6.2.)
Home screen to Main Menu
1 Select “Cal. Condition” with the [ ↑ ] [ ↓ ] keys, and press the
[Enter/Menu] key.
2 Select “N” with the [ ↑ ] [ ↓ ] keys, and press the [Enter/Menu] key.
3 Select “Opt Length” with the [ ↑ ] [ ↓ ] keys, and press the [Enter/Menu]
key.
Calibration Menu screen
Calibration Condition Setup screen
Number of Sampling
Lengths Setup screen