EasyManua.ls Logo

Mitutoyo SJ-210 - Measurement

Mitutoyo SJ-210
407 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
No. 99MBB122A
4-4
4.3 Measurement
To start measurement, set the SJ-210 on a workpiece and press the [START/STOP] key.
While measurement is being performed, the measured profile is displayed. After
measurement has been completed, the measurement result is displayed for confirmation.
4.3.1 Setting the workpiece and SJ-210
Setting the workpiece and SJ-210
When the workpiece surface is larger than the SJ-210, place the SJ-210 on the workpiece.
For surface roughness measurement to be successful, it should be performed on a firm
base that is insulated as well as possible from all sources of vibration. When
measurement is performed being subject to significant vibrations, results may be
unreliable.
TIP
In cases where the measured surface is smaller than the SJ-210 or where the surface is
curved (cylindrical, etc.), install the SJ-210 using an appropriate optional accessory. For
information about optional accessories, refer to Chapter 14, “INSTALLING THE SJ-210
WITH OPTIONAL ACCESSORIES”.
1 Position the work piece so that the measured surface is level.
2 Place the SJ-210 on the workpiece.
In this operation, support the SJ-210 by reference surfaces A and B on the bottom of
the driving unit, as shown below.
Setting the SJ-210 on the workpiece
A
B

Table of Contents

Related product manuals