10. OPERATING ENVIRONMENT SETUP
No. 99MBB122A
10-47
10.16.1 Items restored to original values when restoring factory default settings
Measurement data: all data is cleared.
Measurement conditions, parameter detail settings, GO/NG judgment result tolerance
values
Measurement conditions
Standard Profile Parameter Filters λc λs
Number of
sampling lengths
Pre-travel
Post-travel
Traversal
speed
Range
ISO1997 R
3
(Ra, Rq, Rz)
GAUSS 0.8 0.25 5 ON 0.5 AUTO
Parameter detail settings
Parameter Definition Unit
Number of
sections
Height of the
slice level
Slice level
Reference
line
Slice depth
Sm/Pc/Ppi/Rc Zp/Zv %
㧙
10.0
㧙 㧙 㧙
HSC Peak %
㧙
10.0
㧙 㧙 㧙
mr N
㧙
1
㧙 㧙
0%
0.1
m (3.9 in)
mr(c) Peak % 2
㧙
10%, 15%
㧙 㧙
σc
㧙 㧙
1
㧙
25% 10%
㧙
AnnexA ON
㧙 㧙 㧙 㧙 㧙 㧙
GO/NG judgment: the mean and tolerance values are all 0.
Calibration measurement nominal values, calibration conditions, calibration history
(except for the last calibration performed)
Nominal value: 2.95 (standard type, retractable type), 1.00 (transverse tracing type)
Calibration history: to be cleared.
Calibration conditions (standard type, retractable type)
Standard Filters λc Number of sampling lengths Traversal speed Range
JIS1994 GAUSS 2.5 5 0.75 AUTO
Calibration conditions (transverse tracing type)
Standard Filters λc Number of sampling lengths Traversal speed Range
JIS1994 GAUSS 0.8 5 0.5 AUTO