No. 99MBB122A
8-24
8.4.4 Setting calculation conditions when mr[c] (tp for ANSI) is selected
When the parameter “mr(c)” (“tp” for ANSI) is selected, the slice level must also be set as a
calculation condition.
Operating procedure (Refer to “ Accessing the Submenu screen” in Section 8.1.)
Home screen to Main Menu
1 Select “DetailsSetting” with the [ ] [ ] keys, and press the
[Enter/Menu] key.
2 Select “mr(c)” (“tp” for ANSI) with the [ ] [ ] keys, and press the
[Enter/Menu] key.
3 Select the reference for the slice level, and press the [Enter/Menu] key.
“Peak”: Set from the highest peak of the evaluation profile
“Base”: Set from the mean line of the evaluation profile
The background of the selected slice level reference turns blue.
Submenu screen
Setting Details Selection screen
mr(c) Setup screen
mr(c) Setup screen