No. 99MBB122A
18-16
18.4 Traversal Length
In the SJ-210, the traversal length is the sum of the measured length, the approach travel
length, the pre-travel length, and the post-travel length.
NOTE
• The pre-travel length and the post-travel length vary depending on the filter used.
When the pre-travel and post-travel settings are set to NO, the traversal length is
reduced by the pre-travel length and the post-travel length.
For details about enabling/disabling the pre-travel and post-travel, refer to 7.9, “Setting
Pre-travel/Post-travel”.
Measuring operation
Measurement starts from the origin position. When measurement has been completed,
the detector returns to the origin.
■ Traversal length
When the 2CR filter is selected
Traversal length (When 2CR filter is selected)
When GAUSS filter is selected
Traversal length (When GAUSS filter is selected)
Data from the pre-travel length and the post-travel length are calculated assuming that
their lengths are ℓ/2.
1 Cycle Reciprocal motion 1 mm/s (0.02 in/s)
Approach
travel
0.5 mm (0.02 in)
Pre-travel
ℓ
Evaluation length
ℓ
n
ℓ=F
Traversal length
Normal measuring position
Origin
Approach
travel
0.5 mm (0.02 in)
Pre-travel
ℓ/2
Evaluation length
ℓ
n
ℓ=F
Traversal length
Normal measuring position
Origin
Post-travel
ℓ/2