No. 99MBB122A
6-16
6.5.2 Modifying the roughness standard
Set the roughness standard according to the precision roughness specimen.
NOTE
• Be careful when modifying the roughness standard, as the profile filter may be
automatically modified as a result.
■ Operating procedure (Refer to “■ Accessing the Calibration Menu screen” in Section 6.2.)
Home screen to Main Menu
1 Select “Cal. Condition” with the [ ↑ ] [ ↓ ] keys, and press the
[Enter/Menu] key.
2 Select “Standard” with the [ ↑ ] [ ↓ ] keys, and press the [Enter/Menu]
key.
3 Select roughness standard which is compatible with the precision
roughness specimen with the [ ↑ ] [ ↓ ] keys, and press the
[Enter/Menu] key.
Roughness Standard Setup screen
Calibration Menu screen
Calibration Condition Setup screen