No. 99MBB122A
18-10
18.2.2 Filters
■ Types of filters
The following 3 types of filters are available.
Filter Amplitude characteristics Phase characteristics Amplitude transmission
at the cut-off value
2CR 2CR Without phase correction 75%
PC75 2CR Phase correction 75%
GAUSS Gaussian Phase correction 50%
The characteristics of each filter are explained below.
The attenuation characteristic of each filter is represented by the characteristics of a
high-pass filter.
2CR
This filter has the same attenuation characteristic as 2 C-R circuits that are connected
in series and that have identical time constants.
The attenuation characteristic is -12 dB/oct, and the amplitude transmission at the
cut-off value is 75%, as shown in figure below.
Attenuation characteristic of 2CR filter
Attenuation characteristic:
2
3
1
1
c
H
Am
litude transmission
%
Am
litude transmission
dB