7.8 Setting the Evaluation Length to an Arbitrary
Length
The SJ-210 is capable of setting the evaluation length to an arbitrary length in a range
between 0.30 mm to 16.00 mm (0.0118 in to 0.6299 in).
The evaluation length is the distance of the arbitrary length minus the pre-travel/post-travel
lengths.
When pre-travel/post-travel is set to “OFF”, the evaluation distance is equal to the arbitrary
length.
Setting an arbitrary length and the traversal length/evaluation length
NOTE
The setting range of an arbitrary evaluation length depends on the cutoff value and filter
settings. When performing measurement at an arbitrary evaluation length, set the length
after setting a cut-off value and a filter.
Note that the procedure for setting the evaluation length to an arbitrary length differs
when “R-Motif” is selected as the evaluation profile. For information about the setting
procedure, refer to “ Operating procedure (when a Motif evaluation profile (R-Motif) is
specified)” below.
TIP
For more information about the relationship between the evaluation profile and
pre-travel/post-travel, refer to 18.4, “Traversal Length”.
When pre-travel/post-travel is set to “OFF”, pre-travel/post-travel is calculated with
overlapping data included.
Approach travel
0.5 mm (0.019 in)
Pre-
travel
Evaluation length
Traversal length
Normal measuring position
Origin
Post-
travel
Arbitrary length