6. CALIBRATION
No. 99MBB122A
6-27
6.7 Setting the Stylus Alarm
Stylus alarm is a function to notify when to replace the detector or perform regular
calibration by setting the threshold for cumulative measured value.
This section explains how to specify the threshold.
NOTE
Be aware that the cumulative distance is completely cleared, when power supply from
built-in battery is cut or “RestToDefault” in the “Set Environ.” is performed.
Operating procedure (Refer to “ Accessing the Calibration Menu screen” in Section 6.2.)
Home screen to Main Menu
1 Select “Stylus Alarm” with the [ ] [ ] keys, and press the
[Enter/Menu] key.
2 Press the “Thresh” ([Red] key).
TIP
To clear the cumulative distance, press the “ClearDis”
([Blue] key) on the Stylus Alarm screen.
3 Specify the threshold.
TIP
The value is set to 0 when the “AC” ([Blue] key) is pressed.
For information about numeric value entry, refer to 2.5,
“Entering Numeric Values/Characters”.
Calibration Menu screen
Stylus Alarm screen
Threshold Setup screen