6. CALIBRATION
No. 99MBB122A
6-27
6.7 Setting the Stylus Alarm
Stylus alarm is a function to notify when to replace the detector or perform regular
calibration by setting the threshold for cumulative measured value.
This section explains how to specify the threshold.
NOTE
• Be aware that the cumulative distance is completely cleared, when power supply from
built-in battery is cut or “RestToDefault” in the “Set Environ.” is performed.
■ Operating procedure (Refer to “■ Accessing the Calibration Menu screen” in Section 6.2.)
Home screen to Main Menu
1 Select “Stylus Alarm” with the [ ↑ ] [ ↓ ] keys, and press the
[Enter/Menu] key.
2 Press the “Thresh” ([Red] key).
TIP
• To clear the cumulative distance, press the “ClearDis”
([Blue] key) on the Stylus Alarm screen.
3 Specify the threshold.
TIP
• The value is set to 0 when the “AC” ([Blue] key) is pressed.
• For information about numeric value entry, refer to 2.5,
“Entering Numeric Values/Characters”.
Calibration Menu screen
Stylus Alarm screen
Threshold Setup screen