No. 99MBB122A
8-18
8.4.2 Setting calculation conditions when HSC is selected
When the “HSC” parameter is selected, the height of the count level, a calculation condition,
must be set.
Operating procedure (Refer to “ Accessing the Submenu screen” in Section 8.1.)
Home screen to Main Menu
1 Select “DetailsSetting” with the [ ] [ ] keys, and press the
[Enter/Menu] key.
2 Select “HSC” with the [ ] [ ] keys, and press the [Enter/Menu] key.
3 Select the reference for the height of the count level, and press the
[Enter/Menu] key.
“Peak”: Set from the highest peak of the evaluation profile
“Base”: Set from the mean line of the evaluation profile
The background of the selected reference of the height of the count
level turns blue.
Submenu screen
Setting Details Selection screen
HSC Setup screen
HSC Setup screen