7.10 Modifying the Traversing Speed
The traversing speed can be modified according to settings such as cut-off value (λc) and
the upper limit length.
The cutoff value (sampling length) and the traversing speed
The traversing speed can be set according to the cutoff value (λc) and the upper limit
length as detailed in the table below.
Cutoff value
(sampling
length)
mm (in)
A mm (in)
(for R-Motif)
Traversing speed
mm/s (in/s)
0.08 (0.003)
0.25, 0.5
(0.010, 0.020)
0.25 (0.01) 0.02 (0.001) 0.25, 0.5
(0.010, 0.020)
0.8 (0.03) 0.10 (0.004)
0.25, 0.5
(0.010, 0.020)
2.5 (0.1) 0.5 (0.020) 0.25, 0.5, 0.75
(0.010, 0.020, 0.030)