8. MODIFYING PARAMETERS
No. 99MBB122A
8-3
8.2 Selecting the Displayed Parameters (Parameter
Customization)
Parameter customization functions can be used to set what is calculated and displayed.
8.2.1 Customizing parameters
Overview of the parameter customization function
The instrument is initially set at the factory to calculate and display the most commonly
used parameters. For other parameters, you can use the parameter customization function
to specify their calculation and display settings.
By calculating and displaying only specified parameters, the time required to calculate
measurement results is shortened, and the key operations for switching parameter display,
etc. can be simplified.
Also, parameters can be selected or deselected all at once.
TIP
The definition of each parameter is given in 18.5, “Definitions of the SJ-210 Roughness
Parameters”.
When the Sm, Pc, or Ppi parameter is selected, the height of the count level must also
be set. Refer to 8.4.1, “Setting calculation conditions when Sm, Pc, Ppi or Rc is
selected” for setting procedure details.
When the HSC parameter is selected, the height of the count level must also be set.
Refer to 8.4.2, “Setting calculation conditions when HSC is selected” for setting
procedure details.
When the mr parameter is selected, the number of sections, reference line, and slice
level must also be set. Refer to 8.4.3, “Setting calculation conditions when mr is
selected” for setting procedure details.
When the mr[c] parameter is selected, the slice level must also be set. Refer to 8.4.4,
“Setting calculation conditions when mr[c] (tp for ANSI) is selected” for setting
procedure details.
• When the δc parameter is selected, the reference line and slice level must also be set.
Refer to 8.4.5, “Setting calculation conditions when δc (Htp for ANSI) is selected” for
setting procedure details.