6. CALIBRATION
No. 99MBB122A
6-25
6.5.8 Modifying the measuring range
Set the measuring range according to the precision roughness specimen.
■ Operating procedure (Refer to “■ Accessing the Calibration Menu screen” in Section 6.2.)
Home screen to Main Menu
1 Select “Cal. Condition” with the [ ↑ ] [ ↓ ] keys, and press the
[Enter/Menu] key.
2 Select “Range” with the [ ↑ ] [ ↓ ] keys, and press the [Enter/Menu] key.
3 Select measuring range which is compatible with the precision
roughness specimen with the [ ↑ ] [ ↓ ] keys, and press the
[Enter/Menu] key.
Selected measuring range appears on the Calibration Condition
Setup screen.
TIP
• Press the [Esc/Guide] key to return to the previous screen.
Calibration Menu screen
Calibration Condition Setup screen
Measurement Range Setup screen
Calibration Condition Setup screen