6. CALIBRATION
No. 99MBB122A
6-11
6.4 Setting the Nominal Value of the Precision
Roughness Specimen
Set the nominal value according to the precision roughness specimen.
TIP
The nominal value to be set is the Ra value of which the precision roughness specimen
has been calibrated.
Operating procedure (Refer to “ Accessing the Calibration Menu screen” in Section 6.2.)
Home screen to Main Menu
1 Select “Nom Val.” with the [ ] [ ] keys, and press the [Enter/Menu]
key.
2 Set the nominal value.
TIP
Pressing the “AC” ([Blue] key) sets the value to 0.
To change the position of decimal point, place the cursor at
the desired position and press the “Decimal” ([Red] key).
For information about numeric value entry, refer to 2.5,
“Entering Numeric Values/Characters”.
3 Press the [Enter/Menu] key to accept the “Nom Val.”.
TIP
To cancel settings input, press the [Esc/Guide] key instead
of the [Enter/Menu] key.
Calibration Menu screen
Nominal Value Setup screen