10.5 Calibrating Drive Unit Speed and Settings
Other than the standard drive unit, the SJ-210 also supports the detector retracting type
drive unit as well as the transverse tracing type drive unit. As specifications such as
start-up distance and maximum traversal distance differ depending on the drive unit used,
the drive unit must be set up.
IMPORTANT
• When the drive unit is exchanged, traversing speed calibration must be performed.
There is a possibility that calculation results may be affected.
This is an explanation of the drive unit settings on the display unit.
TIP
• For information about changing the drive unit refer to 3.2, “Attaching and Detaching the
Drive/Detector Unit”.
• To perform traversing speed calibration, the instrument must be calibrated using the
included roughness specimen.
For the placement of the roughness specimen and the SJ-210, refer to 6.1, “Calibration
Preparation”.
■ Operating procedure (Refer to “■ Accessing the Operating Environment Setup Menu screen” in Section
10.1.)
Home screen to Main Menu
1 Select “Drive” with the [ ↑ ] [ ↓ ] keys, and press the [Enter/Menu] key.
2 Select a drive unit type with the [ ↑ ] [ ↓ ] keys, and press the
[Enter/Menu] key.
Operating Environment
Setup menu screen
Drive Unit Setup screen