10.5 Calibrating Drive Unit Speed and Settings
Other than the standard drive unit, the SJ-210 also supports the detector retracting type
drive unit as well as the transverse tracing type drive unit. As specifications such as
start-up distance and maximum traversal distance differ depending on the drive unit used,
the drive unit must be set up.
IMPORTANT
When the drive unit is exchanged, traversing speed calibration must be performed.
There is a possibility that calculation results may be affected.
This is an explanation of the drive unit settings on the display unit.
TIP
For information about changing the drive unit refer to 3.2, “Attaching and Detaching the
Drive/Detector Unit”.
• To perform traversing speed calibration, the instrument must be calibrated using the
included roughness specimen.
For the placement of the roughness specimen and the SJ-210, refer to 6.1, “Calibration
Preparation”.
Operating procedure (Refer to “ Accessing the Operating Environment Setup Menu screen” in Section
10.1.)
Home screen to Main Menu
1 Select “Drive” with the [ ] [ ] keys, and press the [Enter/Menu] key.
2 Select a drive unit type with the [ ] [ ] keys, and press the
[Enter/Menu] key.
Operating Environment
Setup menu screen
Drive Unit Setup screen