No. 99MBB122A
13-2
13.1 SPC Data Output
By connecting the SJ-210 to a DP-1VR Digimatic data processor (optional accessory) with
an SPC cable (optional accessory), calculation results are output using SPC and can be
statistically processed and printed. Apart from recent measurements, data saved to the
memory card can be loaded and output with SPC for statistical processing and printing.
IMPORTANT
• Only the calculation results of parameters with the SPC mark ( ) can be output as
SPC data. Parameter names, etc., are not output.
• When outputting parameter calculation results for statistical processing, take care to not
include data obtained with differing parameters.
An error may occur when multiple pieces of parameter data with differing units and
decimal place positions are output to the Digimatic data processor.
The operation flow of outputting SPC data is described below.
Connect the DP-1VR
→ 13.1.1, “Connecting the SJ-210 and the DP-1VR
”
Executing the actual measurement of the
roughness specimen
→ Chapter 4, “MEASUREMENT OPERATION”
Set calculation result output to “SPC”
→ 10.3.1, “Setting data output to SPC”
Selecting parameters to output
→ 13.1.2, “Selecting Parameters”
Outputting calculation results
→ 13.1.3, “Outputting SPC Data”
Loading saved data for output
→ 9.4, “Loading Measurement Data”
When outputting
When outputting saved data