3. SETTING UP THE SJ-210
No. 99MBB122A
3-13
3.4 Power Supply
A built-in battery and an AC adapter are provided to supply power to the SJ-210.
When a built-in battery is used, the SJ-210 itself can turn the power on without connecting
the AC adapter to the SJ-210.
When an external power supply is available, connect the AC adapter to the SJ-210 and
turn on the power.
IMPORTANT
• Upon purchase the built-in battery switch is set to OFF. Be sure to set the built-in battery
switch to ON before using this instrument.
• When the AC adapter is connected while the built-in battery switch is set to OFF, the
below icon is displayed. Disconnect the AC adapter, set the built-in battery switch to ON,
and then reconnect the AC adapter.
The icon when the built-in battery switch is OFF.
• When the built-in battery power is almost consumed, the power cannot be turned on.
Charge the built-in battery to drive the SJ-210 again with the battery. However, notice
that the measurement conditions and results saved in the built-in memory are cleared.
• When the built-in battery switch is OFF, the measurement results and conditions are
cleared. Keep the built-in battery switch ON unless the SJ-210 is not used for a long
period of time (more than 2 to 3 weeks).
• The following items are saved in the SJ-210's internal memory even when the built-in
battery switch is OFF or when the internal battery is replaced.
• Detector calibration factor
• Drive unit traversing speed calibration factor
• Drive unit type
• Language
• Unit
• Decimal Point
• Date Format