6. CALIBRATION
No. 99MBB122A
6-21
6.5.5 Modifying the number of sampling lengths (N)
Set the number of sampling lengths (N) according to the precision roughness specimen.
■ Operating procedure (Refer to “■ Accessing the Calibration Menu screen” in Section 6.2.)
Home screen to Main Menu
1 Select “Cal. Condition” with the [ ↑ ] [ ↓ ] keys, and press the
[Enter/Menu] key.
2 Select “N” with the [ ↑ ] [ ↓ ] keys, and press the [Enter/Menu] key.
3 Select number of sampling lengths which is compatible with the
precision roughness specimen with the [ ↑ ] [ ↓ ] keys, and press the
[Enter/Menu] key.
Selected number of sampling lengths appears on the Calibration
Condition Setup screen.
TIP
• Press the [Esc/Guide] key to return to the previous screen.
Calibration Menu screen
Calibration Condition Setup screen
Number of Sampling Lengths
Setup screen
Calibration Condition Setup screen