7. MODIFYING MEASUREMENT CONDITIONS
No. 99MBB122A
7-25
7.12 Reprocessing Calculation Results
After measurement, measurement conditions can be changed, and the result is
recalculated.
The SJ-210 has a function to recalculate the measured data after performing a roughness
measurement by modifying the measurement conditions. When this recalculation function
is on, measurement data is recalculated and displayed based on the modified
measurement conditions.
■ Measurement conditions that can be modified for recalculation
The SJ-210 can perform recalculation after the modification of the following measurement
conditions.
Roughness standard Evaluation profile
Filter Number of sampling lengths (reduction)
Parameters GO/NG judgment
NOTE
• When the cut-off value or arbitrary length is modified and the sample pitch and data
point conditions do not match, recalculation may not be possible.
• The recalculation function cannot be used when the number of sampling lengths has
been increased, for example from “1” to “3”.
• When pre-travel/post-travel is set to “ON” from “OFF”, recalculation may be unavailable.
• When the filter or evaluation profile is modified and pre-travel/post-travel conditions do
not match, recalculation may not be possible.