6.5.3 Modifying profile filters
Set the profile filters according to the precision roughness specimen. Profile filters can be
modified when multiple profile filters which are compatible with roughness standard exist.
SJ-210 modifies the profile filter automatically according to the roughness standard, when
roughness standard is modified.
■ Operating procedure (Refer to “■ Accessing the Calibration Menu screen” in Section 6.2.)
Home screen to Main Menu
1 Select “Cal. Condition” with the [ ↑ ] [ ↓ ] keys, and press the
[Enter/Menu] key.
2 Select “Filter” with the [ ↑ ] [ ↓ ] keys, and press the [Enter/Menu] key.
3 Select profile filter which is compatible with the precision roughness
specimen with the [ ↑ ] [ ↓ ] keys, and press the [Enter/Menu] key.
Calibration Menu screen
Calibration Condition Setup screen
Filter Setup screen