17. PRODUCT SPECIFICATIONS
No. 99MBB122A
17-5
17.3.6 Measurement range and resolution
Measurement range Resolution
Auto depending on the measurement range
0.0016 μm to 0.0256 μm (0.0630 μin to 1 μin)
360 μm (14400 μin) 0.0256 μm (1 μin)
100 μm (4000 μin) 0.0064 μm (0.25 μin)
25 μm (1000 μin) 0.0016 μm (0.0630 μin)
17.3.7 Traversal length
Conditions Pre-travel/post travel length Remark
When P (Primary profile)
and Motif are selected
Pre-travel length = 0 mm (0 in),
Post-travel length = 0 mm (0 in)
When the R
(Roughness) and the
2CR are selected
Pre-travel length = λc,
Post-travel length = 0mm (0 in)
When the R
(Roughness) and the
PC75 are selected
Pre-travel length = λc,
Post-travel length = λc
When the R
(Roughness), the
GAUSS and the DF are
selected
Pre-travel length = λc/2,
Post-travel length = λc/2
Approach length (approx. 0.5 mm/0.02 in) and
λs with pre-travel/post-travel length