6.5 Setting Calibration Conditions
Set the following calibration conditions according to the precision roughness specimen.
IMPORTANT
• The default values of the calibration conditions are in accordance with Mitutoyo's
precision roughness specimen. Unless otherwise required, perform calibration
measurement with the default conditions.
Number of measurements per charge
Roughness standard
Filters
Cut-off value (λc)
Number of sampling lengths (N), or evaluation length (arbitrary length)
Traversal speed
Measurement range
Calibration measurement settings are specified on the Calibration Condition Setup screen.
NOTE
• To reset the calibration conditions collectively to the default values (factory settings),
press the “Initial” ([Red] key) on the Calibration Condition Setup screen.