10. OPERATING ENVIRONMENT SETUP
No. 99MBB122A
10-21
3 Set the nominal value for traversing speed calibration.
a On the Calibration Setup screen, press the “Nom Val.” ([Red] key).
NOTE
• Use the included roughness specimen for calibration.
Confirm the placement of the drive unit with the roughness
specimen.
TIP
• To cancel calibration, press the [Esc/Guide] key. Return to
the Operating Environment Setup menu.
b Input the nominal value.
IMPORTANT
• The nominal value must be set to 100m (3937 in)
when using the included roughness specimen.
TIP
• Pressing the “AC” ([Blue] key) sets the value to 0.
To change the position of decimal point, place the cursor at
the desired position and press the “Decimal” ([Red] key).
• For information about numeric value entry, refer to 2.5,
“Entering Numeric Values/Characters”.
c Press the [Enter/Menu] key.
The entered nominal value is displayed on the Calibration Setup
screen.
4 Press the [START/STOP] key to begin measurement.
After measurement the pitch result is displayed.
To cancel the displayed result, press the “Cancel” ([Blue] key).
Nominal Value Setup screen
Calibration Setup screen
Calibration Setup screen
Calibration Setup screen