No. 99MBB122A
10-22
5 Three measurements need to be made, from 0.25mm/s to 0.75
mm/s (0.010 in/s to 0.030 in/s).
6 Press the “Update” ([Red] key).
The traversing speed of the calibration result is changed.
7 Press the [Enter/Menu] key.
TIP
Press the [Esc/Guide] key to return to the previous screen.
Calibration Setup screen