6. CALIBRATION
No. 99MBB122A
6-23
4 Set the arbitrary evaluation length, according to the precision
roughness specimen.
TIP
• Pressing the “AC” ([Blue] key) sets the value to 0.
To change the position of decimal point, place the cursor at
the desired position and press the “Decimal” ([Red] key).
• For information about numeric value entry, refer to 2.5,
“Entering Numeric Values/Characters”.
5 Press the [Enter/Menu] key to accept entered values.
The set arbitrary evaluation length appears on the Calibration
Condition Setup screen.
TIP
• Press the [Esc/Guide] key to return to the previous screen.
Arbitrary Length Setup screen
Calibration Condition Setup screen