17. PRODUCT SPECIFICATIONS
No. 99MBB122A
17-13
■ Command
● Control command
Control command basic configuration
Header
(2 bytes)
Sub-field
(3 bytes) Data
*1
EM
(1 byte)
CT * * * * * * * - CR
*1: Data section can be omitted.
Control command
Sub-field Data Meaning
STA None Start measurement/Interrupt processing while
measurement in progress
OFF 00 - 02 (2 bytes) Power off/Setting the auto-sleep function
ESP None Detector retraction
RTN None Reposition the detector to the start position.
STA command
[START/STOP] button operation and start/abort measurement are performed.
* Measurement is aborted when this command is issued during measurement.
Command
Header Sub-field EM
CT STA CR
Response (normal)
Header EM
OK CR
Response (abnormal)
Header Sub-field EM Meaning
NG * * * CR * * *: Refer to “● Error codes”.