No. 99MBB122A
17-18
CON command
Reads measurement/evaluating conditions. Shares a common format with the write
command.
Command
Header Sub-field EM
RD CON CR
Response
Header Data EM
OK * * * * * CR
Data Bytes: number of bytes from the tip of data
Bytes Settings Description
0
* (standard) 0: JIS1982, 1: JIS1994, 2: JIS2001, 3: ISO1997, 4: ANSI,
5: VDA, 6: Free
1 * (Profile) 0: P, 1: R, 2: DF, 3: R-MOTIF
2
* (Cutoff length λc) 0: 0.08, 1: 0.25, 2: 0.8, 3: 2.5
λs is set according to λc.
3 * * (Number of sampling lengths) 00 - 10
5
* *. * * (Arbitrary evaluation length) 0.10 - 16.00 (When the number of sampling length is 00)
Unit [mm]
10
* (Upper limit of motif length A)
1: 0.02, 2: 0.1, 3: 0.5
Upper limit of motif length B is set according to A.
11
㧖 Filters 0:2CR75ޔ1:PC75ޔ2:GAUSSޔ3:None
PAR command
Reads number of parameters currently customized.
Command
Header Sub-field EM
RD PAR CR
Response
Header Data EM
OK * * CR
* *Number of pieces