Remote Commands
R&S
®
FSW
1319User Manual 1173.9411.02 ─ 43
OFF | 0
Switches the coupling off
ON | 1
Switches the coupling on
*RST: 1
Example:
INST:COUP:USER3 'Spectrum1','All
Windows','Attenuation','Spectrum 2','All
Windows','Attenuation',BID,ON
Synchronizes the attenuation between the channels named
'Spectrum1' and 'Spectrum 2' in both directions and turns on the
coupling.
Manual operation: See "Edit coupling definition" on page 723
INSTrument:COUPle:USER<uc>:CHANnel:LIST?
This command queries the names of the measurement channels that can be synchron-
ized.
Suffix:
<uc>
.
Index of a user-defined parameter coupling. Which indexes are
available is defined by INSTrument:COUPle:USER<uc>:
NUMBers:LIST?.
Return values:
<SynchronizableChannel>Comma-separated list of strings
All channels that can be synchronized.
Example:
INST:COUP:USER:CHAN:LIST?
Result:
'SPEC1','AD1','All Spectrum','All
Channels','All Analog Demod'
Usage: Query only
Manual operation: See "Channel 1 / Channel 2" on page 725
INSTrument:COUPle:USER<uc>:ELEMent:LIST? [<ChannelName>, <Parameter>]
This command queries parameters that can be synchronized.
Suffix:
<uc>
.
irrelevant
Query parameters:
<ChannelName> Optional SCPI parameter.
String containing the name of a measurement channel.
Configuring the R&S FSW