Remote Commands
R&S
®
FSW
1323User Manual 1173.9411.02 ─ 43
Example:
INST:COUP:USER:NEW? 'All Spectrum','All
Windows','Attenuation','','','',BID,ON
Result:
3
Synchronizes the attenuation between all Spectrum channels in
both directions and turns on the coupling. Also returns the index
number of the user-defined coupling.
Usage: Query only
Manual operation: See "Parameter 1 / Parameter 2" on page 724
See "Add New User Coupling" on page 724
INSTrument:COUPle:USER<uc>:NUMBers:LIST?
This command queries the index numbers of user-defined parameter couplings. The
index numbers are used to refer to the specific coupling in remote commands with a
USER<uc> suffix.
Suffix:
<uc>
.
irrelevant
Return values:
<Index> Comma-separated list of strings
Index numbers of all available user-defined couplings
Note that the returned index numbers are not necessarily the
same as those shown in the user interface.
Example:
INST:COUP:USER:NUMB:LIST?
Result:
'1','2','4'
Number '3' is not returned, because a coupling with that index
does not exist anymore.
Usage: Query only
Manual operation: See "Index" on page 723
INSTrument:COUPle:USER<uc>:RELation <Direction>
This command selects the direction in which synchronization works.
Note that the command is not available if you synchronize over all channels or all
channels of the same application.
Suffix:
<uc>
.
Index of a user-defined parameter coupling. Which indexes are
available is defined by INSTrument:COUPle:USER<uc>:
NUMBers:LIST?.
Parameters:
<Direction> LTOR | RTOL | BIDir
Configuring the R&S FSW