Common Analysis and Display Functions
R&S
®
FSW
526User Manual 1173.9411.02 ─ 43
For EMI measurements, these functions are not available; use Automatic Peak Search
instead (see Chapter 7.13.4.2, "EMI Final Measurement Configuration", on page 346).
Remote command:
CALCulate<n>:MARKer<m>:MAXimum:AUTO on page 1162
CALCulate<n>:MARKer<m>:MINimum:AUTO on page 1164
Search Limits
The search results can be restricted by limiting the search area or adding search con-
ditions.
Search Limits (Left / Right) ← Search Limits
If activated, limit lines are defined and displayed for the search. Only results within the
limited search range are considered.
For details on limit lines for searches, see "Peak search limits" on page 548.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe] on page 1159
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT on page 1160
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt on page 1160
Search Threshold ← Search Limits
Defines an absolute threshold as an additional condition for the peak search. Only
peaks that exceed the threshold are detected.
Remote command:
CALCulate<n>:THReshold on page 1161
Use Zoom Limits ← Search Limits
If activated, the peak search is restricted to the active zoom area defined for a single
zoom.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits:ZOOM[:STATe] on page 1161
Deactivating All Search Limits ← Search Limits
Deactivates the search range limits.
Remote command:
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe] on page 1159
CALCulate<n>:THReshold:STATe on page 1161
9.3.3.2 Marker Search Settings for Spectrograms
Access: "Overview" > "Analysis" > "Markers" > "Search"
or: [MKR TO] > "Search Config"
Spectrograms show not only the current sweep results, but also the sweep history.
Thus, when searching for peaks, you must define the search settings within a single
time frame (x-direction) and within several time frames (y-direction).
These settings are only available for spectrogram displays.
Marker Usage

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals