Remote Commands
R&S
®
FSW
964User Manual 1173.9411.02 ─ 43
Usage: Event
14.5.8.5 Performing a Spurious Measurement
The following commands are required to perform a Spurious measurement:
SENS:SWE:MODE LIST, see [SENSe:]SWEep:MODE on page 919
INITiate<n>[:IMMediate] on page 836, see Chapter 14.5.1, "Performing Mea-
surements", on page 834
14.5.8.6 Retrieving and Saving Settings and Results
The following commands analyze and retrieve measurement results for Spurious mea-
surements.
Useful commands for spurious emission measurements described elsewhere
●
CALCulate<n>:LIMit<li>:FAIL? on page 1224
●
TRACe<n>[:DATA] on page 1143
●
TRACe<n>[:DATA]:MEMory? on page 1144
●
TRACe<n>[:DATA]:X? on page 1145
14.5.8.7 Programming Example: Spurious Emissions Measurement
In the following example, the Spurious Emissions measurement is configured by defin-
ing ranges and parameters to create the following sweep list.
Configuring and Performing Measurements