6-8 Inlets and probes
3. Evaporate any solvent used in loading by placing the sample cup(s) in a
warm location.
Fitting the sample cup to the probe tip
To fit the sample cup to the probe tip
Use tweezers to fit the sample cup to the sample cup holder in the end of the
probe tip.
Tip: To ensure good thermal contact with the heater, fit the sample cup firmly
in the holder.
Warning: The probe components can be contaminated with toxic
materials. Wear nitrile gloves at all times while handling the
components.
Caution: To avoid contaminating the sample, after flame-cleaning,
keep the sample cup free of grease (for example, from laboratory
surfaces) and handle with tweezers.
Caution: When fitting the sample cup to the sample cup holder, ensure
that it is firmly held by the sample cup holder; if the sample cup is
loose, it can become dislodged when inserting the probe in the ion
source.
Sample cup
Sample cup holder