3-12 Scanning techniques
B/E (product ion) function
Use the B/E Scan function editor is used to set up scans in which the magnet’s
field strength (B) is scanned (as for a Magnet Scan), but with the addition of a
scan of ESA1 and ESA2. The electric fields (E) are scanned such that the ratio
B/E remains constant. This function detects only the product ions of a specific
precursor ion.
B/E Scan function editor: