7-56 Routine maintenance
Disassembling the re-entrant coupling mechanism
To disassemble the re-entrant coupling mechanism
1. Slide the peg out of the mounting bracket.
2. Remove the two screws that secure the tee-block to the U-bracket.
3. Remove the nut that secures the peg to the U-bracket.
4. Remove the peg from the U-bracket.
Disassembling the probe contact assembly
Probe contact assembly components
A
B
C
D
D
D
E
F
G
H